Marine Deep-Water Multi-Parameter Water Quality Sensor IS844D

Multi-parameter water quality sensor IS844D is specifically designed for demanding marine and deep-water conditions, providing accurate monitoring up to 200 meters deep, allows simultaneous measurement of multiple water parameters, including pH, ORP, turbidity, dissolved oxygen, etc. The multi-parameter water quality sensor IS844D features automatic probe cleaning, fast response algorithm (<30s), digital RS485 Modbus output and efficient alarm system, simplifying long-term unattended water quality monitoring. Suitable for data acquisition platforms and automatic water quality stations.
  • Designed for deep-sea and marine environments, supporting depths up to 200 meters (pH/ORP probes up to 100 meters).
  • Rugged pressure-resistant structure with custom titanium alloy watertight connectors, ensuring durability in challenging conditions.
  • Measures up to 7 parameters simultaneously with up to 6 interchangeable digital probes, including pH, ORP, conductivity, turbidity, dissolved oxygen, blue-green algae, chlorophyll, and oil-in-water.
  • Digital RS485 output and standard Modbus protocol for easy integration into data acquisition platforms and automatic monitoring stations.
  • Equipped with an automatic cleaning device, automatic cleaning at regular intervals to eliminate probe errors caused by pollution or air bubbles, ensuring long-term reliable monitoring.
  • Each single-parameter probe can be replaced, calibrated, and maintained independently in the field, providing users with great flexibility and convenience.
  • High-efficiency acquisition algorithm ensures fast response within 30 seconds, and alerts users to abnormal events such as downtime, communication failures, and cleaning requirements.
ParameterSpecification
Measuring Depth0 to 200 m
InterfaceSupports up to 6 freely configurable sensor, 1 removable central cleaning brush
DimensionsΦ84 mm × 476 mm
Operating Temperature-5 to 50 ℃
Calibration DataStored in the single-parameter probe; supports independable calibration in the field
OutputRS-485 output, MODBUS protocol
Brush ControlDefault cleaning every 30 minutes (interval adjustable)
Power Consumption0.8 W (non wiping), 2 W (wiping)
Power Supply RequirementsDC 12–24 V, ≥ 1 A
Protection ClassIP68
Weight3.25 kg (main unit + 6 sensors)
Status AlarmsInternal power anomaly alarm, internal communication anomaly alarm, cleaning brush anomaly alarm
Cable LengthWith waterproof connector, 50 m (default), customizable
MaterialsPVC, titanium alloy, anti-fouling copper

 

Sensor TypeMeasurement PrincipleMeasuring RangeAccuracyResolution
Dissolved OxygenOptical Method0 to 20 mg/L or 0 to 200 % saturation±1 % or ±0.3 mg/L, whichever is greater0.01 mg/L
Turbidity90° Scattering Light0 to 1000 NTU (optional 0 to 4000 NTU)±5 % or 0.3 NTU, whichever is greater0.01 NTU
Conductivity (Salinity)Four-electrodeConductivity: 0 to 200 mS/cm
Salinity: 0 to 175 ppt
TDS: 0 to 128000 mg/L
±1 % or 0.01 mS/cm, whichever is greater
±1 ppt
/
0.01 mS/cm
0.01 ppt
/
Oil-in-WaterUV Fluorescence Method0 to 50 ppm (optional 0 to 150 ppm)Linearity: R² > 0.9990.01 ppm
pHGlass Electrode Method0 to 14 pH±0.02 pH0.01
ORPPlatinum Electrode Method-1999 to 1999 mV±20 mV1 mV
ChlorophyllOptical Method0 to 500 μg/L±5 % or ±0.5 μg/L, whichever is greater0.01 μg/L
CyanobacteriaOptical Method0 to 300,000 cells/mL (optional 0 to 2 million)Linearity: R² > 0.9991 cells/mL
DepthDifferential Pressure0 m to 2 m…200 m H₂O< 0.1 % F.S.0.01 m

Specifications are subject to change without prior notice.