Benchtop conductivity meter with Large Display and Data Storage for Beginners

The L111C07 benchtop conductivity meter for water is designed for basic laboratory and educational use with a large LCD screen, flexible electrode holder, dual power supply method. Featuring 256 data storage sets, 1 to 3 point manual calibration, auto temperature compensation, EC, TDS, salinity simultaneous measurement, the L111C07 benchtop conductivity meter simplifies routine water quality testing and delivers reliable performance.
  • A entry-level benchtop EC meter with large easy-to-read LCD screen display and easy operation function.
  • Featuring removable electrode holders that can be mounted on either side, supports to set up three electrodes simultaneously.
  • Single-tap switch between conductivity, TDS, and salinity.
  • Quick and easy 1 to 3 points manual calibration, with calibration guide and standard solution recognition, ensuring accuracy and reliability.
  • With 256 groups of data storage, making measurements track simple and efficient.
  • Conductivity electrode constant, reference temperature, temperature compensation coefficient, TDS conversion factor, are adjustable depending on various measurement conditions.
  • User-friendly operation functions include auto power off, automatic temperature compensation, and auto reading hold.
  • The standard k=1 plastic graphite conductivity/temp. electrode offers strong corrosion resistance and easy maintenance.
  • Powered by either alkaline batteries or adapter, providing flexibility for various working conditions.
ParameterSpecification
Conductivity Measuring Range0 to 400 mS/cm (auto-range):
0.00 to 19.99 μS/cm;
20.0 to 199.9 μS/cm;
200 to 1999 μS/cm;
2.00 to 19.99 mS/cm;
20.0 to 199.9 mS/cm
TDS Measuring Range0 to 400 g/L
Salinity Measuring Range0 to 260 ppt
Resistivity Measuring Range0 to 100 MΩ·cm
Resolution0.01 / 0.1 / 1 μS/cm; 0.01 / 0.1 mS/cm
Accuracy±1.0% F.S ±1 digit
Temperature Compensation Range0 to 50°C (manual or automatic)
Cell Constant (K)0.01 / 0.1 / 1 / 10 cm⁻¹
Temperature Measuring Range0 to 100°C
Temperature Resolution0.1°C
Temperature Accuracy±0.5°C ±1 digit
Data Storage256 sets
Power SupplyAAA batteries × 4 or 110–240V power adapter

Specifications are subject to change without prior notice.